跳至主要内容

IEC60112 Tracking Index Test Apparatus for PTI and CTI Testing

 https://www.pegotester.com/IEC60112_tracking_index_tester.html#_jcp=3_17

1. Introduction

The tracking index test apparatus is designed according to IEC60112, UL746A, GB.T4207, GB4706.1 and TMD 3628-92. On the surface of the solid insulating material, a certain voltage is applied between the platinum electrodes (2 mm × 5 mm) , and drops the pollution liquid (0.1% NH4CL) with specified droplet volume in fixed height (35mm) for 30s to evaluate the resistance of the surface of solid insulating materials under the combined action of electric field and contaminated medium, and its Comparative tracking index (CTI) and proof tracking index (PTI) are determined. Tracking index test apparatus is mainly used in the industries of lighting appliance, low voltage appliance, household appliance, motor, electronic appliance, instrument and etc. It is a good tool for research department, production department and QC department.

The tracking index test apparatus is controlled by single chip and push buttons. The test apparatus is built-in air exhaust, the front panel is transparent for easy observation of the experimental process.

 

Proof Tracking Index (PTI): the voltage value of the material subjected to 50 drops of electrolyte without electrical tracking, the unit is volt.

Comparative Tracking Index (CTI): The highest voltage value at which the surface of the material can withstand 50 drops of electrolyte without forming electrical tracking, the unit is volt.

Test solutions: The preparation ratio is 1:1000, that is 0.1g of NH4CI with 100g (100ml) of distilled water

 

2. Parameter:

Electrode material

platinum (Pt)  purity 99.8%

Electrode size

2mm±0.1mm)×(5mm±0.1mm)×(40mm±5mm), length≥12mm, Angle of chisel-edged: 30°±2°

Electrode distance

4.0mm±0.01mm, angle: 60°±5°

Electrode pressure

1.00±0.05N

Electrode distance

4.0mm ± 0.01mm

Angle

60°±5

Electrode voltage

100~600V

Height of the drops

35mm±5mm adjustable

Size of drops

20 drops:  0.380g~0.480g, 50 drops: 0.997g~1.147g

Interval of the drops

30s±0.1s, 50 drops: 24.5min±2min

Test solutions resistance

Solution A: 0.1%NH4Cl3.95±0.05Ωm, solution B: 1.7±0.05Ωm

Number of drops

0~9999 (presettable, record and display)

Test wind speed

0.2m/s

Short circuit current drop

1.0A±0.1A drop 8%

Leakage indicator

0.50A±10% , 2.00s±10%

Thermal radiation baffle

Automatic masking

Safe operation interface

automatically power off upon open the glass window

Standard

IEC60112, UL746A, ASTM D 3638-92, DIN 53480

Working studio

0.5m3 ( customizable 0.1m3, 0.5m3 or 0.75m3 according to different standard), exhaust hole: Φ100mm

Outer dimension

600mm (W)*500mm (D)*1000mm (H)

Working Power

220V/50-60Hz, 0.6KVA

 

评论

此博客中的热门博文

JIS D0203 rain nozzle and spray nozzle test chamber

https://www.pegotester.com/JISD0203_rain_and_spray_test_chamber.html The spray nozzle and rain nozzle test chamber is designed to meet the test requirements of JISD0203 R1/R2 rain test and S1/S2 spray test. The rain nozzle has 2 holes and spray nozzle has 40 holes. Type Symbol Test Purpose Test Method Rain Test R1 Auto parts against water direct contact water pressure: 0.01Mpa, water flow: 1.9L/min, water temperature: ordinary temperature, test time: 10min R2 Auto parts against rain or spray water   indirect contact water pressure: 0.03Mpa, water flow: 3.2L/min, water temperature: ordinary temperature, test time: 10min Spray Test S1 Auto parts against rain or jet water direct contact water pressure: 0.10Mpa, water flow: 24.5L/min, water temperature: ordinary temperature, test time: 30min S2 Auto parts against riptide water pressure: 0.30Mpa, water flow: 39.2L/min, water temperature: ordinary temperature, test time: 1h  

UL746A High Current ARC Ignition Test Apparatus 32.5A/COSφ0.5

UL746A High Current ARC Ignition Test Apparatus 32.5A/COSφ0.5   1.  Introduction: The high current arc ignition tes t apparatus is designed according to  IEC60950-1:2001  clause 4.7.3.6 , BS/EN60950-1:2001, BS7002:2001, GB4943-2001 and UL746A, IEC60947.1:1999, GB/T14048.1-2000, IEC730.1:1993, GB14536.1-1998 and etc..  It ’ s used to measure and assess  solid insulating material against to resistance to ignition from arcing electrical sources. A power 240V(AC)/60Hz is to be supplied to the test electrodes  (fixed electrode and movable electrode) , a series (inductive-resistive) air core impedance is to be provided to yield a short circuit current of 32.5A and a power factor of 0. 5. The sample is be to positioned with the electrodes making initial contact on the surface of the sample, the circuit is to be energized and the cyclic arcing started. The timing of arcs is to be adjusted to a rate of 40 arcs/min, the rate of electrode separation is to be 254±...

what equipment can be used for IPX3 and IPX4 water ingress protection test?

1. Spray nozzle 1.  Introduction The spray nozzle hand-held device is designed according to IEC60529 figure 5, it is it is apply to verify the protection against spraying water (IPX3) and splashing water (IPX4) for large size sample. The hand-held device is consisted by spray nozzle, moving shield, counterweight, hose and pressure gauge. The spray nozzle is made of brass and has 121 holes with diameter 0.5mm (1 hole at center, 1 inner circles of 12 holes at 30 0  pitch, 4 outer circles of 24 holes at 15 0 pitch). The moving shield is made of aluminum, the counterbalanced shield is in place for IPX3 test, the water pressure is adjusted to give the specified delivery rate, the pressure to achieve this delivery rate will be in the range of 50kPa to 150kPa, and it should be kept constant during the test, the test duration is 1min/m 2  of the calculated surface of the enclosure and with a minimum duration of 5 min. For IPX4 test, the counterbalanced shield is removed fro...